![]() ![]() | DIGITAL MANOMETER (ACCURACY 0,1%)LEO 2 has two keys. The left key turns the instrument on, to select the functions and the pressure units. The right key is used to switch between the MAX.- and MIN.-value. RESET: With the RESET-function, the MAX.- and MIN.-value is set to the actual pressure value. ZERO: The ZERO-function allows to set any value as a new Zero reference. Barometric pressure variations can thus be compensated. The factory setting of the Zero for the ranges -1 3 bar or -1 30 bar is at 0 bar absolute. For sealed gauge pressure measurements, activate ZERO SEt at ambient pressure. Instruments with the ranges 0 300 bar or 0 700 bar are calibrated in a sealed gauge mode with 1 bar absolute pressure as a Zero reference. CONT: The instrument turns off 15 Min. after the last key function. Activating CONT (Continuous) deactivates this automatic turn-off. UNITS: All standard instruments are calibrated in bar. The pressure can be indicated in the the following units: bar, mbar/hPa, kPa, MPa, PSI, kp/cm2. LEO 2 has 7/16 UNF male thread and is delivered complete with an O-ring seal swivel fitting which allows LEO 2 to be rotated through 360°. Only minimal tightness is needed to seal up to 300 bar. Process pressure connection is G 1/4 male. Download to view Specs |
| | |