![]() ![]() | DIGITAL MANOMETER (PEAK MEASURING RATE: 5 kHz / ACCURACY: 0,2%)The LEO 1 has two modes; MANO-Mode: The pressure is measured twice per second and displayed. The top display indicates the actual pressure, the bottom display shows the MAX.- or MIN- pressure since the last RESET. PEAK-Mode: The pressure is measured 5000 times per second and is brought to the display twice per second. The top display shows the actual pressure, the bottom display the peak- (MAX.) or trough-pressure (MIN.). LEO 1 has two keys. The left key turns the instrument on, to to select the functions and the pressure units. The right key is used to switch between the MAX.- and MIN.-value in both modes. RESET: With the RESET-function, the MAX.- and MIN.-value is set to the actual pressure value. ZERO: The ZERO-function allows to set any value as a new Zero reference. Barometric pressure variations can thus be compensated. The factory setting of the Zero for the ranges -1 3 bar or -1 30 bar is at 0 bar absolute. For sealed gauge pressure measurements, activate ZERO SEt at ambient pressure. Instruments with the ranges 0 300 bar, 0 700 bar or 0 1000 bar are calibrated in a sealed gauge mode with 1 bar absolute pressure as a Zero reference. CONT: The instrument turns off 15 Min. after the last key function. Activating CONT (Continuous) deactivates this automatic turn-off. UNITS: All standard instruments are calibrated in bar. The pressure can be indicated in the following units: bar, mbar/hPa, kPa, MPa, PSI, kp/cm2 Optional Accessories: Protective rubber covering and carrying bag. Download to view Specs |
| | |
| Copyright Wales Instruments & Wales LEO 1 / LEO 1 Ei |